Continuous glucose monitors are sold over the counter now, which means a lot of people without diabetes are wearing one for two weeks to see what bread does to them. The hardware is genuinely good. What it measures is narrower than the app implies, and the gap between the two is where most of the conclusions go wrong.
The sensor is a filament a few millimeters long sitting under the skin, and it is not in your blood. It reads interstitial fluid, the liquid between cells, using an enzyme that reacts with glucose and produces a current the transmitter converts to a number. Glucose arrives in that fluid after it arrives in the blood, so the reading trails by roughly ten to fifteen minutes. On a flat afternoon that lag is invisible. During a sharp rise it means the number on your phone is a recent past, not a present, and the peak you see is lower and later than the one in your bloodstream.
Accuracy is usually quoted as MARD, the mean absolute relative difference against a lab measurement. Current sensors land somewhere around 8 to 10 percent. That is fine for watching a trend and loose for a single reading: a displayed 100 can sit within a band of roughly 90 to 110 and still be performing to spec. Two readings a few points apart are the same reading.
There are known artifacts worth recognizing before you interpret anything. A sensor needs a warm-up period of an hour or more after insertion, and the first day often reads differently from the rest. Lying on the arm wearing it can produce a compression low, a dip that is pressure on the filament and not a drop in glucose. Signal loss and a flat line mean the transmitter lost contact, not that your body did something interesting. And a new sensor on day one against an old one on day fourteen will disagree, which is a property of the sensors rather than a change in you.
Then the normal physiology, which the apps are getting better about but still frame as a problem. Glucose rises after you eat. That is the system working. A rise to the 140s after a real meal and a return toward baseline inside two hours is ordinary in a person without diabetes. Exercise can push glucose up during the effort through adrenaline and down afterward. Poor sleep, a cold, caffeine and stress all move the line without any food involved. A score that treats every excursion as a failure is grading you on being alive.
Where the device earns its money is comparison, and only within itself. Same breakfast, two preparations, same time of day, same sleep: the difference between those two curves is real information about you, and no population study can give it to you. Oatmeal that spikes you and not your partner is a fact about your own digestion. That is the experiment the hardware is good at.
What it cannot do is diagnose anything. An over-the-counter monitor is not a screening tool, the thresholds that define prediabetes are defined on blood tests rather than interstitial readings, and a scary-looking fortnight of data is a reason to get an A1C and a fasting glucose from a lab, not a diagnosis on its own. Running the numbers in the other direction, toward reassurance, is no safer: a flat two weeks does not rule anything out either.
On cost, these run roughly $50 to $100 a month depending on the brand and whether you buy sensors singly or on subscription, with each sensor lasting about two weeks. For most curious people the honest plan is one or two sensors, run a few deliberate comparisons, write down what you learned, and stop. The marginal value of month six is close to zero unless a clinician asked you to keep going.
The short version: it measures interstitial glucose on a ten to fifteen minute delay with about a ten percent error band, which is excellent for comparing your own meals and useless as a verdict on your health.